Page 1 - OPERATION MANUAL

OPERATION MANUAL DAKOTA ULTRASONICS DDFFXX--88 SSeerriieess Ultrasonic Flaw Detectors (Manual 2 of 2) P/N P-250-0004

Page 2

Dakota Ultrasonics 6 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multip

Page 3

Dakota Ultrasonics 96 11.7 Zero The DFX-8 has a feature to manually adjust the electronic zero. This feature used in conjunction with manually adjust

Page 4

DFX-8 Series Ultrasonic Flaw Detectors 97 6) Press the LEFT and RIGHT arrow keys multiple times using the left button wheel to scroll the digit loca

Page 5

Dakota Ultrasonics 98 2) Use the UP and DOWN arrow keys multiple times on the left button wheel to scroll through the sub menu items until BRIGHTNES

Page 6 - 1.2 Electrical Warning

DFX-8 Series Ultrasonic Flaw Detectors 99 3) Press the ENTER key to display a list of the color scheme options. 4) Press the UP and DOWN arrow keys

Page 7 - 2.1 Gauge Type

Dakota Ultrasonics 100 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key mul

Page 8

DFX-8 Series Ultrasonic Flaw Detectors 101 Filled – Draws a filled version of the waveform. The procedure to select the look and feel

Page 9 - Load Default Setup

Dakota Ultrasonics 102 Alternatively, press the LEFT and RIGHT arrow keys multiple times using the left button wheel. 2) Use the UP and DOWN arrow

Page 10

DFX-8 Series Ultrasonic Flaw Detectors 103 None Line Dots

Page 11 - CHAPTER THREE

Dakota Ultrasonics 104 the DISP menu is highlighted and displaying the submenu items. Alternatively, press the LEFT and RIGHT arrow keys multiple t

Page 12

DFX-8 Series Ultrasonic Flaw Detectors 105 Immersion Testing Considerations:  There is a significant difference between sound traveling in water

Page 13 - 3.3 CAL – Menu

7 CHAPTER THREE KEYBOARD, MENU, & CONNECTOR REFERENCE 3.1 Menu Key (Operation & Sub Menus) The Menu key activates the primary menu

Page 14

Dakota Ultrasonics 106 The procedures for enabling and adjusting the IM measurement mode are outlined below: Activating the (IM) Mode 1) Press the M

Page 15 - 3.6 GT1 – Menu

DFX-8 Series Ultrasonic Flaw Detectors 107 11.15 Filters (Wide & Narrow Bands) The DFX-8 includes a wide band filter with a range of 1.8MHz to 19M

Page 16 - 3.7 GT2 – Menu

Dakota Ultrasonics 108 11.16 Gain Step Size Adjusting Gain Step Size 1) Press the MENU key once to activate the menu items tab. Press the MENU key

Page 17 - 3.9 AWS – Menu

DFX-8 Series Ultrasonic Flaw Detectors 109 8) Press the OK key to set the gain step and return to the menu screen, or ESC to cancel entering the gain

Page 18 - 3.11 DAC – Menu

Dakota Ultrasonics 110 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key mul

Page 19 - 3.12 DGS – Menu

DFX-8 Series Ultrasonic Flaw Detectors 111 The diagram above is an RF view, showing both the positive and negative going cycles for purposes of expl

Page 20 - 3.13 SETUP – Menu

Dakota Ultrasonics 112 Note: The screenshots above show where all the possible locations for polarity selection are found in the DFX-8. For more inf

Page 21 - 3.14 DATA – Menu

DFX-8 Series Ultrasonic Flaw Detectors 113 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right

Page 22 - 3.17 CLR (clear) Key

Dakota Ultrasonics 114 Activating and Using Compare 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times

Page 23 - 3.22 ENTER key

DFX-8 Series Ultrasonic Flaw Detectors 115 The DFX-8+ is also equipped with an analog outputs to send a 0-10 volt electrical signal to an external dev

Page 24

Dakota Ultrasonics 8 Activating and Navigation of the Menu Items 1) Press the MENU key once to activate the menu items tab. Press the MENU key multi

Page 25 - 3.29 DFX-8 Overview

Dakota Ultrasonics 116 The DFX-8 is equipped with an internal clock to time and date stamp the log, setup and screen capture files for reporting/docum

Page 26

DFX-8 Series Ultrasonic Flaw Detectors 117 5) Press the LEFT and RIGHT arrow keys multiple times on the right button wheel to scroll the value. 6) R

Page 27 - Navigating Hot Menus

Dakota Ultrasonics 118 3) Use the LEFT and RIGHT arrow keys multiple times using the left button wheel to scroll the KEY CLICK options. 4) Once the

Page 28

DFX-8 Series Ultrasonic Flaw Detectors 119 2) Use the UP and DOWN arrow keys multiple times using the left button wheel to scroll through the sub me

Page 29

Dakota Ultrasonics 120 Note: At any point during a PEAK scan the operator may opt to clear the wave drawing and start over. 1) Press the CLEAR key at

Page 30

DFX-8 Series Ultrasonic Flaw Detectors 121 Note: If the CAPTURE feature has been activated, pressing the FREEZE key a second time will capture the sc

Page 31 - Display View

122 CHAPTER TWELVE DATA STORAGE – SETUP, EDIT, & VIEW FILES 12.1 Introduction to Grid and Sequential file formats The DFX-8 is equipped with two d

Page 32

DFX-8 Series Ultrasonic Flaw Detectors 123 Grid File Formats Sequential Log Formats Important Note: For the duration of this ch

Page 33 - Setting Range

Dakota Ultrasonics 124 Grid/Seq Log Name : Can contain a combination of up to 20 numeric, alpha, or special characters listed in the first section of

Page 34

DFX-8 Series Ultrasonic Flaw Detectors 125 Grid Log Sequential Log 4) Press the ENTER key to display the new Grid or

Page 35 - 4.3 Gain

DFX-8 Series Ultrasonic Flaw Detectors 9 3.2 Probe – Menu TYPE: Enables the user to select the type of transducer being used, either Single

Page 36

Dakota Ultrasonics 126 9) Use the CLR key to backspace if necessary. 10) Repeat steps 6 - 9 until the Grid or Seq Name (File Name) is completed. 11)

Page 37 - 4.4 Gates

DFX-8 Series Ultrasonic Flaw Detectors 127 3) Use the UP, DOWN, LEFT, & RIGHT arrow keys on the left button wheel to highlight the appropriate

Page 38 - Activating a Gate

Dakota Ultrasonics 128 Sequential: The sequential file format can be viewed as a file as a single column of up to 512 possible rows (readings), and a

Page 39 - LEFT and

DFX-8 Series Ultrasonic Flaw Detectors 129 Grid Log Sequential Log 3) GRID LOG: Use the LEFT, & RIGHT arrow ke

Page 40

Dakota Ultrasonics 130 Grid Log Sequential Log 1) Use the UP and DOWN arrow keys on the left button wheel to scroll

Page 41

DFX-8 Series Ultrasonic Flaw Detectors 131 SEQ LOG: Use the UP, DOWN, LEFT, & RIGHT arrow keys on the left button wheel to highlight the approp

Page 42 - Dakota Ultrasonics

Dakota Ultrasonics 132 Selecting the Auto Increment Direction The Auto Increment feature gives the user the ability to specify which direction to adv

Page 43 - DFX-8 Setup

DFX-8 Series Ultrasonic Flaw Detectors 133 Saving a Grid Once all the parameters are set, the user has the option of saving or canceling the new grid

Page 44

Dakota Ultrasonics 134 Grid Log Sequential Log 3) Press the OK key to save the New Grid or Seq Log, or the ESC key to

Page 45

DFX-8 Series Ultrasonic Flaw Detectors 135 Key Feature: When the DFX-8 is displaying a “grid log”, pressing the OK key will initiate an advance to ro

Page 46

Dakota Ultrasonics 10 MATL 2PT: Performs a two-point calibration. This option allows the user to automatically calculate the velocity by entering a

Page 47

Dakota Ultrasonics 136 12.4 Viewing stored readings It is sometimes necessary to go back and view the stored readings and B-Scans using the DFX-8 with

Page 48

DFX-8 Series Ultrasonic Flaw Detectors 137 4) The user may opt to clear a specific reading and save a new one at any time. Press the CLR key in the

Page 49

Dakota Ultrasonics 138 3) Press the ENTER key to display the File List Box. 4) Use the UP and DOWN arrow keys on the left button wheel to scroll

Page 50

DFX-8 Series Ultrasonic Flaw Detectors 139 3) Press the ENTER key to activate the confirmation screen. 4) Press the OK key to delete All Files from m

Page 51

Dakota Ultrasonics 140 12.6 Editing a Grid (File) Once a grid has been created and saved to memory, the user can edit the Name, Comments, or Increment

Page 52

DFX-8 Series Ultrasonic Flaw Detectors 141 5) Press the ENTER key to activate the Alpha Edit box – Only used when editing the NOTE. 6) Use the UP,

Page 53

Dakota Ultrasonics 142 11) Press the OK key to save the changes or the ESC key to cancel editing the file parameters. 12) Press the MEAS key to r

Page 54 - 5.3 Angle Beam Calibration

DFX-8 Series Ultrasonic Flaw Detectors 143 the DATA menu is highlighted and displaying the submenu items. Alternatively, press the LEFT and RIGHT a

Page 55

Dakota Ultrasonics 144 7) Press the MEAS key to return to the measure screen. 12.8 Closing an active File - Close A user might not have a current req

Page 56 - Selecting the Material Type

DFX-8 Series Ultrasonic Flaw Detectors 145 3) Press the ENTER key to close the active file. Note: Following the key press, the CLOSE text will be

Page 57

DFX-8 Series Ultrasonic Flaw Detectors 11 3.5 TUNE – Menu MEASURE MODE: This menu item is only included in our DFX-8+. Toggles a variet

Page 58 - 6.2 Additional Comments

146 CHAPTER THIRTEEN SETUPS – CREATE, STORE, EDIT, & RECALL 13.1 Introduction to Setups Often times, users are faced with a variety of tasks and a

Page 59 - 6.3 Setup Trigonometry Mode

DFX-8 Series Ultrasonic Flaw Detectors 147 Opening a Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple t

Page 60

Dakota Ultrasonics 148 5) Press the ENTER key to activate the confirmation screen. 6) Press the OK key to load the setup from memory. 7) Press the ME

Page 61

DFX-8 Series Ultrasonic Flaw Detectors 149 Saving a Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple tim

Page 62

Dakota Ultrasonics 150 5) When the parameter to edit is highlighted, press the ENTER key to activate the Alpha Edit Box. 6) Use the

Page 63

DFX-8 Series Ultrasonic Flaw Detectors 151 11) Use the UP and DOWN arrow keys on the left button wheel to scroll through the setups until the targ

Page 64 - 7.1 Introduction to AWS

Dakota Ultrasonics 152 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key mul

Page 65 - Activating & Setting AWS

DFX-8 Series Ultrasonic Flaw Detectors 153 6) Press the OK key to delete the Setup File. 7) Finally, press the MEAS key to return to the measurement

Page 66

Dakota Ultrasonics 154 Using the Default Setup 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab r

Page 67

DFX-8 Series Ultrasonic Flaw Detectors 155 13.6 Selecting a Language The DFX-8 is equipped with a language option. Currently, only a few languages su

Page 68

Dakota Ultrasonics 12 GATE1: Gates allow the user to view a specific inspection/measurement range, or specific regions of the waveform, while ignorin

Page 69

156 CHAPTER FOURTEEN SOFTWARE, FILE TRANSFER, & UPGRADES 14.1 Computer System Requirements DakView will run on many different operating systems:

Page 70 - 8.2 Additional Comments

DFX-8 Series Ultrasonic Flaw Detectors 157 14.4 Using the XFER menu (DFX-8) The XFER menu of the DFX-8 will be used in conjunction with the DakView PC

Page 71 - 8.3 Creating a DAC Curve

Dakota Ultrasonics 158 Selecting Storage Device 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to

Page 72

DFX-8 Series Ultrasonic Flaw Detectors 159 Copying Files Between Storage Devices 1) Press the MENU key once to activate the menu items tab. Press t

Page 73 - Drawing the DAC

Dakota Ultrasonics 160 5) Press the ENTER key to either transfer ALL files, or display the Individual List of files. 6) Use the UP and DOWN arrow

Page 74

DFX-8 Series Ultrasonic Flaw Detectors 161 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab ri

Page 75

162 APPENDIX A - VELOCITY TABLE Material L-Wave velocity in/µs L-Wave velocity m/s S-Wave Velocity in/µs S-Wave Velocity in/µs Aluminum 0.2510 63

Page 76 - 9.3 Creating a TCG Curve

DFX-8 Series Ultrasonic Flaw Detectors 163 Steel Stainless" 0.2230 5664 .1230 3120 Teflon 0.0540 1372 .2500 6350 Tin 0.1310 3327 .0660 1670 Tit

Page 77

164 APPENDIX B - SETUP LIBRARY Num Name Comment 1 Gn/AGC Velocity 1 Enter Custom Name 2 … 3 … 4 … 5 … 6 … … …

Page 78 - Drawing the TCG

165 WARRANTY INFORMATION  Warranty Statement  Dakota Ultrasonics warrants the DFX-8 against defects in materials and workmanship for a period of t

Page 79

DFX-8 Series Ultrasonic Flaw Detectors 13 POLARITY2: The gates can be set to detect on either the positive or negative polarity, or stroke, of the ha

Page 81 - DISTANCE, GAIN, SIZE (DGS)

PCSoftware&FirmwareNote:Allsoftwaredownloadsandupgradesareavailableonourwebsiteatnoadditionalcharge.Take

Page 82 - 10.2 DGS Diagram

Dakota Ultrasonics 14 REFERENCE GAIN: The gain (in dB) required to set the indication level to a specific percentage of full screen height (FSH). Th

Page 83 - EFF = Effective Diameter

DFX-8 Series Ultrasonic Flaw Detectors 15 MEAS: Toggles the units displayed as a relationship of either the signal amplitude and DAC curve, or just t

Page 85 - 10.5 Attenuation Parameters

Dakota Ultrasonics 16 DVK: Used to correct for increases in gain based on a curved reference standard, to an actual back wall reflector. Refer to pa

Page 86 - Setting DGS Parameters

DFX-8 Series Ultrasonic Flaw Detectors 17 3.14 DATA – Menu NEW: Allows the user the ability to create a new alpha numeric grid, or sequentia

Page 87

Dakota Ultrasonics 18 ANALOG OUTPUT: DFX-8+ only. Proportional outputs (amplitude or distance), 0-10 volts. Speed is determined by the PRF setting.

Page 88

DFX-8 Series Ultrasonic Flaw Detectors 19 The primary functions of the CLR key, is to clear a measurement from a grid or sequential log files cell loc

Page 89 - Drawing the DGS Curve

Dakota Ultrasonics 20 The ENTER key is used in the overall menu selection process, to activate list and edit boxes, display and save measurements/wave

Page 90

DFX-8 Series Ultrasonic Flaw Detectors 21 The ON/OFF key simply powers the unit either ON or OFF. Note: Unit will automatically power off when idle

Page 91

Dakota Ultrasonics 22 iv. DAC – Distance Amplitude Correction v. DGS – Distance Gain Size  View i. RF: Radio Frequency. ii. R+: Rectified posit

Page 92 - Selecting the Pulse Width

DFX-8 Series Ultrasonic Flaw Detectors 23 Always Displayed & Active Displayed Only When Active The diagram above is a screenshot of al

Page 93

Dakota Ultrasonics 24 arrow keys to scroll through the Hot Menu items. Alternatively, the MEAS & ESC keys can be used to scroll through the Hot M

Page 94

DFX-8 Series Ultrasonic Flaw Detectors 25 USB Type B Connector Refer to Diagram: The USB connector is a common type B connector found on a variety of

Page 95

CHAPTER ONE INTRODUCTION ... 1CHAPTER TWO QUICK START GUIDE ...

Page 96 - Changing the Pulser Voltage

Dakota Ultrasonics 26 Analog Output (LEMO 1 – 5 pin) Refer to Diagram: The DFX-8+ is also equipped with analog outputs 0-10volt, and is update

Page 97

27 CHAPTER FOUR SETTING UP FOR MEASUREMENT 4.1 Selecting Display View The DFX-8 includes a variety of viewing options for the waveform display. There

Page 98 - Adjusting Velocity

Dakota Ultrasonics 28 using the left button wheel. For quick access and editing, refer to navigating the Hot Menus page 22. 2) Use the UP and DOWN

Page 99

DFX-8 Series Ultrasonic Flaw Detectors 29 Setting Delay 1) Using the left button wheel, press the LEFT and RIGHT arrow keys to scroll through the

Page 100 - Adjusting Zero

Dakota Ultrasonics 30 Setting Range 1) Using the left button wheel, press the LEFT and RIGHT arrow keys to scroll through the 9 individual Hot Me

Page 101 - Adjusting Brightness

DFX-8 Series Ultrasonic Flaw Detectors 31 4.3 Gain The gain feature of the DFX-8 is used to adjust the signal amplitude to a predetermined reference h

Page 102 - Selecting a Color Scheme

Dakota Ultrasonics 32 2) Additionally, the increment GAIN STEP size can be adjusted at any time by pressing the GAIN button and displaying the Dig

Page 103 - Setting a DIM time

DFX-8 Series Ultrasonic Flaw Detectors 33 2) Use the UP and DOWN arrow keys on the left button wheel to scroll through the sub menu items until GA

Page 104

Dakota Ultrasonics 34 detect on reflections that occur inside the boundaries of the gate. Anything outside of the start and range/width of the gate w

Page 105 - Setting the Look & Feel

DFX-8 Series Ultrasonic Flaw Detectors 35 until the GT1 menu is highlighted and displaying the submenu items. Alternatively, press the LEFT and RIG

Page 107 - Setting Wave Averaging

Dakota Ultrasonics 36 4) Press the LEFT and RIGHT arrow keys multiple times using the left button wheel to scroll the digit locations. 5) Repeat ste

Page 108

DFX-8 Series Ultrasonic Flaw Detectors 37 1) Press the UP and DOWN arrow keys multiple times on the left button wheel until THRESHOLD1 is highlighte

Page 109

Dakota Ultrasonics 38

Page 110

39 CHAPTER FIVE CALIBRATION In this section we’ll focus on calibrating the DFX-8. There are a number of ways to go about performing a calibration, bu

Page 111

Dakota Ultrasonics 40 The gauge will require a few basic adjustments in order to get underway with the calibration process. For this example, in orde

Page 112 - Adjusting Gain Step Size

DFX-8 Series Ultrasonic Flaw Detectors 41 2) Press the ENTER key to display the Digits Edit Box. 3) Press the UP and DOWN arrow keys multiple times

Page 113 - Selecting the Detect

Dakota Ultrasonics 42 Activating a Gate 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right an

Page 114 - 11.18 Polarity

DFX-8 Series Ultrasonic Flaw Detectors 43 3) Press the UP and DOWN arrow keys multiple times on the left button wheel to scroll the highlighted valu

Page 115 -

Dakota Ultrasonics 44 2) Press the ENTER key to display the Digits Edit Box. 3) Press the UP and DOWN arrow keys multiple times on the left button w

Page 116

DFX-8 Series Ultrasonic Flaw Detectors 45 Two Point Calibration – Straight Beam Note: Calibration standards representing the minimum and ma

Page 117 - 11.20 Compare

CHAPTER ONE INTRODUCTION The Dakota Ultrasonics model DFX-8 is both, an ultrasonic flaw detector, as well as an ultrasonic A/B Scan thickness scope

Page 118

Dakota Ultrasonics 46 2) Press the LEFT and RIGHT arrow keys on the right button wheel to move the gate over the waveform to be measured, as demonst

Page 119

DFX-8 Series Ultrasonic Flaw Detectors 47 7) Press the OK key to calculate the velocity and return to the menu screen, or ESC to cancel the one point

Page 120 - Setting Date & Time

Dakota Ultrasonics 48 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multi

Page 121 - 11.23 Key Click

DFX-8 Series Ultrasonic Flaw Detectors 49 Note: A velocity and zero value will be calculated and displayed in the fields a shown above. 8) Finally,

Page 122 - Capturing Screen Shots

Dakota Ultrasonics 50 3) Press the MEAS key once to highlight the WIDTH1 cell. Alternatively, use the UP and DOWN arrow keys on the left button wh

Page 123

DFX-8 Series Ultrasonic Flaw Detectors 51 section 5.1 and 5.2, regarding the procedures to setup and calibrate the DFX-8. The following are a list of

Page 124

Dakota Ultrasonics 52 inches. The operator then positions the transducer on an index point, moves the transducer in both directions forward and backw

Page 125

DFX-8 Series Ultrasonic Flaw Detectors 53 3) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right a

Page 126 - CHAPTER TWELVE

54 CHAPTER SIX TRIGONOMETRY MODE 6.1 Introduction to TRIG Trig mode is most commonly used for inspecting welded joints using an angle beam transducer.

Page 127 - Creating a Name

DFX-8 Series Ultrasonic Flaw Detectors 55 When TRIG mode is enabled, the calculated measurements will be displayed below the waveform display area at

Page 128

1.2 Electrical Warning The DFX-8 series gauges contain high voltage pulsers. It’s recommended that the gauge be powered off prior to connecting or di

Page 129

Dakota Ultrasonics 56 Activating & Setting the Parameters Activate Trig Mode 1) Press the MENU key once to activate the menu items tab. Press t

Page 130 - Creating a Note

DFX-8 Series Ultrasonic Flaw Detectors 57 3) Press the UP and DOWN arrow keys multiple times on the left button wheel to scroll the highlighted va

Page 131

Dakota Ultrasonics 58 and RIGHT arrow keys multiple times using the left button wheel to scroll the value. When the correct X-Offset is displayed, pr

Page 132

DFX-8 Series Ultrasonic Flaw Detectors 59 6) Press the OK key to set the CSC OD and return to the menu screen, or ESC to cancel entering the CSC OD. N

Page 133

60 CHAPTER SEVEN AWS – WELD INSPECTION 7.1 Introduction to AWS AWS provides a toolkit for evaluating defects according to the American Welding Society

Page 134

DFX-8 Series Ultrasonic Flaw Detectors 61 inspection, and should additionally be setup prior to enabling the AWS feature. Refer to Chapter Six for in

Page 135

Dakota Ultrasonics 62 until the AWS menu is highlighted and displaying the submenu items. Alternatively, press the LEFT and RIGHT arrow keys multip

Page 136

DFX-8 Series Ultrasonic Flaw Detectors 63 Setting the Reference Gain 1) Press the UP and DOWN arrow keys multiple times on the left button wheel unt

Page 137 - Saving a Grid

Dakota Ultrasonics 64 4) Press the LEFT and RIGHT arrow keys multiple times on the right button wheel to adjust the size or width of GATE1. Note: W

Page 138 - 12.3 Storing a waveform

DFX-8 Series Ultrasonic Flaw Detectors 65

Page 139 - DFX-8 will automatically

3 CHAPTER TWO QUICK START GUIDE This section will cover a basic quick start guide to initially get up and running with the DFX-8 for demonstration pur

Page 140 - 12.4 Viewing stored readings

66 CHAPTER EIGHT DISTANCE AMPLITUDE CORRECTION (DAC) 8.1 Introduction to DAC A DAC curve is a gate with a variable threshold to compensate for a given

Page 141 - Deleting One File

DFX-8 Series Ultrasonic Flaw Detectors 67 %FSH – The top of the waveform display represents 100% full screen height. Any waveform amplitudes generate

Page 142 - Deleting All Data

Dakota Ultrasonics 68 Alternatively, press the LEFT and RIGHT arrow keys multiple times using the left button wheel. 2) Use the UP and DOWN arrow

Page 143

DFX-8 Series Ultrasonic Flaw Detectors 69 RIGHT arrow keys on the right button wheel to toggle and select which curve or gate to use as the alarm trig

Page 144 - Editing a Grid

Dakota Ultrasonics 70 ex. 1 ex. 2 2) Press the left button wheel, press the LEFT and RIGHT arrow keys to scroll throu

Page 145

DFX-8 Series Ultrasonic Flaw Detectors 71 6) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right,

Page 146

72 CHAPTER NINE TIME CORRECTED GAIN (TCG) 9.1 Introduction to TCG The TCG feature increases gain as distance increases, in order to achieve an overall

Page 147

DFX-8 Series Ultrasonic Flaw Detectors 73 Setting TCG Parameters Activating TCG 1) Press the MENU key once to activate the menu items tab. Press th

Page 148

Dakota Ultrasonics 74 3) Press the UP and DOWN arrow keys multiple times on the left button wheel to scroll the highlighted value. 4) Press the LE

Page 149

DFX-8 Series Ultrasonic Flaw Detectors 75 1) With all the parameters set in the previous section, press the MEAS key once to return to the main measu

Page 150 - 13.2 Opening a Setup

Dakota Ultrasonics 4 Dim: Off Rect Wave: Filled Detect Mark: None Note: The following features are set to off: GT2, TRIG

Page 151

Dakota Ultrasonics 76 Note: You must set 3 points to draw a TCG curve, and a total of 16 points can be set. Any point can be cleared by pressing the

Page 152 - 13.3 Saving a Setup

77 CHAPTER TEN DISTANCE, GAIN, SIZE (DGS) 10.1 Introduction to DGS (AVG) The DGS (Distance, Gain, Size) method was introduced in 1958 as a way to stan

Page 153 - Saving a Setup

Dakota Ultrasonics 78 Considerations:  Should only be used with single element probes.  The effective diameter must be estimated for standard tran

Page 154

DFX-8 Series Ultrasonic Flaw Detectors 79 Where: N = Near Field Length DEFF = Effective Diameter = Wave Length (velocity/frequency) In order to a

Page 155 - Deleting a Setup

Dakota Ultrasonics 80 DEFF = Effective Diameter K = Aspect Ratio A = Largest Dimension 0.97 = Fixed Ratio Calculated results can also be compared to

Page 156

DFX-8 Series Ultrasonic Flaw Detectors 81 Trigger Curve: Enables the user to set the detection curve to either DGS or GATE. This will be the curve u

Page 157 - 13.5 Using the Default Setup

Dakota Ultrasonics 82 normally resulting in greater signal output. DVK is a correction to the gain, in order to simulate the amplitude generated from

Page 158 - Using the Default Setup

DFX-8 Series Ultrasonic Flaw Detectors 83 Alternatively, press the LEFT and RIGHT arrow keys multiple times using the left button wheel. 2) Use the

Page 159

Dakota Ultrasonics 84 Probe Frequency 1) Press the UP and DOWN arrow keys multiple times on the left button wheel until PROBE FREQUENCY is highlight

Page 160 - 14.2 Installing DakView

DFX-8 Series Ultrasonic Flaw Detectors 85 6) Press the OK key to set the ERS and return to the menu screen, or ESC to cancel entering the ERS. Attenua

Page 161

DFX-8 Series Ultrasonic Flaw Detectors 5 >> End TCG DAC DGS SET DATA UTIL XFER TCG DAC DGS OPEN NEW AUTO FIND COPY SETUPS DRAW

Page 162

Dakota Ultrasonics 86 1) Press the UP and DOWN arrow keys multiple times on the left button wheel until REFERENCE GAIN is highlighted, followed by

Page 163

DFX-8 Series Ultrasonic Flaw Detectors 87 Note: The DGS curve uses 16 points to plot the curve. 4) Press the MENU key once to activate the menu items

Page 164

88 CHAPTER ELEVEN ADDITIONAL FEATURES OF THE DFX-8 11.1 Pulse Width The DFX-8 has a square wave pulser with adjustable pulse width option. Pulse widt

Page 165

DFX-8 Series Ultrasonic Flaw Detectors 89 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right a

Page 166 - VELOCITY TABLE

Dakota Ultrasonics 90 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key mult

Page 167

DFX-8 Series Ultrasonic Flaw Detectors 91 Selecting the Damping 1) Press the MENU key once to activate the menu items tab. Press the MENU key multi

Page 168 - SETUP LIBRARY

Dakota Ultrasonics 92 adjusted/optimized, a clean narrow band ultrasonic wave produced, and exceptional output at a given voltage. Therefore, the out

Page 169 - WARRANTY INFORMATION

DFX-8 Series Ultrasonic Flaw Detectors 93 4) Once the desired Pulser Voltage is displayed, press the MEAS key to return to the measurement screen. 1

Page 170

Dakota Ultrasonics 94 3) Press the ENTER key to display the list of material types. 4) Press the UP and DOWN arrow keys multiple times on the left

Page 171 - PCSoftware&Firmware

DFX-8 Series Ultrasonic Flaw Detectors 95 1) Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right